  ¸ðµ¨¸í: ETCR6500
¸ÞÀÌÄ¿: ETCR / China
 
   
 

ETCR6500 High Accuracy Clamp Leaker
 
¥°. Introduction
ETCR6500 High Accuracy Clamp Leaker is specially designed for AC leakage current measurement. It obtains the latest CT and digital integration technology, with characteristics of small volume, high accuracy and perfect function. It is widely used in electric power, telecommunications, meteorology, railways, oil construction, measurement, scientific research and teaching unit, industrial and mining enterprises, etc. Special alloy is used as its sensor core, which applied magnetic shielding technology, results in a almost free interference of external magnetic field.
RS232 interface is available on the meter, The software can achieve online real-time monitoring and historical queries; in dynamic display, with maximum, minimum and mean value indications; also alarm value setting and alarm indication function, historical data reading, saving and print functions, etc. 
The meter can be used for on-line AC current measurement below AC600V, with the functions of peak maintaining, reading holding and data memory. In all, it is high accurate and easy and safe to use.
 

¥±. Technical Specification
Function
AC leakage measurement, AC current measurement, On-line AC current measurement  
Power Supply
6F22 9V  Zi-Mn dry battery 6F22 9V
Process Type
Clamp CT, Integral Approach
Display Type
4 Digits LCD
Clamp Size
35mm¡¿40mm
Meter Dimension
LWH: 175mm¡¿70mm¡¿38mm
LCD Dimension
35mm¡¿21.5mm; Display Dimension: 32mm¡¿15mm
Sampling Rate
2 Times/Second
Frequency
50Hz, 60Hz Automatic identification
Shifting
Automatic Shift
Range
0.01mA¢¦300.0A
Resolution
10uA
Accuracy
(23¡É¡¾3¡É, below 70%RH)
0¢¦60.00A:          ¡¾1.5%¡¾5dgt 
60.00A¢¦199.9A:     ¡¾2%¡¾5dgt   
200.0A¢¦300.0A:     ¡¾4%¡¾5dgt   
Line voltage
Below AC 600V
RS232 Interface
With RS232 interface, Data stored in the memory can be uploaded to the computer, with software monitoring
Data Memory
99 sets, ¡°FULL¡± Blinking when memory is full
Peak Maintaining
Peak value capture hold function, press HOLD without Release to display peak value
Reading Hold
Reading hold function: ¡°DH¡± symbol display
Overflow display
Range overflow function: ¡°OL¡± symbol display
Automatic Shutdown
5 minutes after the boot
Battery Detection
When battery voltage is lower than 7.2V, low battery symbol      will shown to remind battery replacement.
Meter Weight
120g (With battery)
Rated Power
5mW
Working Temperature and Humidity
0¡É¢¦40¡É; below 80%rh
Storage Temperature and Humidity
-10¡É¢¦60¡É; below 70%rh
Insulation Strength
AC2kV/rms (between core and box)
Safety Provisions
IEC1010-1, IEC1010-2-032, Pollution level 2, CAT ¥²(600V)
Accessories
Meter: 1 piece, RS232 Com-cable:1 piece; Software Disk: 1 piece; Battery (6F22 9V): 1 piece
¥². Field Application
 
 