¡ß Inputs |
|
|
Item |
|
Specification |
|
|
|
Input terminal type |
|
|
|
Voltage |
|
Plug-in terminal (safety terminal) |
|
|
|
|
|
|
|
Current |
|
- Direct input: Large binding post - External sensor input: Insulated BNC connector |
|
|
|
Measurement range |
|
|
|
Voltage |
|
- 15 V, 30 V, 60 V, 100 V, 150 V, 300 V, 600 V, 1000 V (for crest factor 3) - 7.5 V, 15 V, 30 V, 50 V, 75 V, 150 V, 300 V, 500 V (for crest factor 6) |
|
|
|
|
|
|
|
Current |
|
- Direct input 500 mA, 1 A, 2 A, 5 A, 10 A, 20 A, 30 A (for crest factor 3) 250 mA, 500 mA, 1 A, 2.5 A, 5 A, 10 A, 15 A (for crest factor 6) - External sensor input 50 mV, 100 mV, 200 mV, 500 mV, 1 V, 2 V, 5 V, 10 V (for crest factor 3) 25 mV, 50 mV, 100 mV, 250 mV, 500 mV, 1 V, 2.5 V, 5 V (for crest factor 6) |
|
|
|
Instrument loss (input resistance) |
|
|
|
Voltage |
|
Approximately 10 M§Ù // 5 pF |
|
|
|
|
|
|
|
Current |
|
- Direct input: Approximately 5.5 m§Ù + approximately 0.03 - External sensor input: Approximately 1 M§Ù |
|
|
|
Instantaneous maximum allowed input (1 second or less) |
|
|
|
Voltage |
|
Peak voltage of 2.5 kV or RMS of 1.5 kV, whichever is lower |
|
|
|
|
|
|
|
Current |
|
Direct input: Peak current of 150 A or RMS of 50 A, whichever |
|
|
|
|
|
|
|
lower |
|
Current sensor input: Peak not to exceed 10 times the range |
|
|
|
Continuous maximum allowed input |
|
|
|
Voltage |
|
Peak voltage of 1.6 kV or RMS of 1.1 kV, whichever is lower |
|
|
|
|
|
|
|
Current |
|
Direct input: Peak current of 90 A or RMS of 33 A, whichever |
|
|
|
|
|
|
|
lower |
|
Current sensor input: Peak not to exceed 5 times the range |
|
|
|
Continuous maximum common mode voltage (50/60 Hz) |
|
|
|
|
|
1000 Vrms |
|
|
|
Influence from common mode voltage |
|
|
|
|
|
Apply 1000 Vrms with the voltage input terminals shorted and the current input terminals open. - 50/60 Hz: ¡¾0.01% of range or less - Reference value up to 200 kHz |
|
|
|
Voltage |
|
¡¾3/range * f% of range or less. However, 3% or less. |
|
|
|
Current direct input and current sensor input: |
|
- ¡¾ (max. range/range)* 0.001 * f% of range or less. - However, 0.01% or less. The units of f are kHz. The maximum rated range within equations is 30 A or 10 V. |
|
|
|
Line filter |
|
Select OFF, 500 Hz, 5.5 kHz, or 50 kHz. |
|
|
|
Frequency filter |
|
Select OFF, or ON |
|
|
|
A/D converter |
|
Simultaneous voltage and current conversion and 16-bit resolution. Conversion speed (sampling rate): Approximately 5 ¥ìsec. See harmonic measurement items for harmonic display. Approximately 10 ¥ìsec for flicker display. |
|
|
|
Range switching |
|
Can be set for each input element. |
|
|
|
Trigger range functions |
|
Increasing range value - When the measured values of U and I exceed 110% of the rating - When the peak value exceeds approximately 330% of the rating (or approximately 660% for crest factor 6)
Decreasing range value - When the measured values of U and I fall to 30% or less range rating, and Upk and Ipk are 300% or less of the lower range value (or 600% for crest factor 6) |
|
|